Specifications |
|
Wavelength range, IR | 7000 to 400 cm-1 |
Wavelength range, NIR | 14000 to 3000 cm-1
determined by detector: 14000-5800 or 9000- 3850 cm-1 |
Resolution | 2 cm-1 |
Interferometer | Pendulum type with roof mirrors |
Output beam diameter | 25 mm |
Optical axis height | 55 mm |
Aperture ratio | f 3.2 |
Beamsplitter, standard IR | Multicoated KBr |
Beamsplitter, option IR | ZnSe |
Beamsplitter, option NIR | Fused Silica |
Frequency reference | VCSEL laser |
IR source | ceramic |
NIR source | Quartz-halogen lamp |
Desiccant possibility | Yes |
Power 12V, 5V, 2.3V; 30W | 12V, 5V, 2.3V; 30W |
Dimensions | W22(26)xD20xH10 cm |
Weight | 4 kg |
Temp. environment | 15 – 28 0C |
Humidity environment | Best below 65% |
General
The INTERSPEC series FTIR modules are stable, rugged units requiring the minimum of external utilities.
For easy access, those system components which have a finite lifetime (i.e. laser, and radiation source) are located at the front side of the module. A yellow LED indicator indicates when the scanning is turned on.
Interspec 407 series modular instruments may be used as an heart of a more complex spectroscopic system. Sampling handling and the whole system is defined by the customer.
Interferometer Performance
The interferometer compartment under cover is 170 x 190 mm wide and 90 mm high. The Interspec 407 series of instruments employ a sealed and desiccated interferometer, ensuring high spectral integrity with low levels of water vapour within the interferometer.
We also offer a version (407Z) that employs a ZnSe moisture insensitive optics to be impervious to water vapour and can be used to advantage in higher humidity environments. Near infrared version (407N) employs a fused silica optics and is insensitive to any influence of water vapour. Resolution in the infrared is available factory default 2 cm-1 . The overall wavelength range is from 7000 to 400cm-1 (IR) or from 14000 to 3000 cm-1 (NIR).
Power supply Unit
407 with power unit
The unit has 12 VDC input and consists of a power supply PCB and motor driver PCB. Communication with the motor PCB via COM port.
Sample holder and Detector Unit Option
Two types of IR or NIR detector units are available: a) with collimating off-axis metal mirrors and b) with off-axis parabolic and off-axis ellipsoidal mirrors and sample slide mount. In the case of IR spectral region, a selected high sensitivity DLATGS pyroelectric detector is available.
In the case of the NIR spectral region, two types of photodiodes are available: Si and InGaAs. The InGaAs detector is with spectral range either from 0.7 to 1.7 µm or from 1.2 to 2.6 µm.
Detector unit with slide sample mount |
Detector unit with parabolic off-axis mirror |
Models:
Interspec 407 – Standard FTIR Module with KBr optics.
Interspec 407Z – Non-hygroscopic FTIR Module with ZnSe optics
Interspec 407N – Near Infra FTIR Module with fused silica optics.